Compact Moisture in Oil Transmitter
EE364
EE364 is a compact and rugged moisture in oil transmitter for inline oil monitoring. It is optimised for high accuracy and excellent long term stability in lubrication, hydraulic and insulation oils, as well as diesel fuel.
EE364 moisture in oil sensor
Water activity (aw) and temperature (t) are accurately measured and the water content (x) in ppm precisely calculated. EE364 features two 4-20 mA outputs and one Modbus RTU interface.
With an optional confugration cable it is possible to perform an adjustment, set the MODBUS interface parameters and configure the analogue outputs.
Typical applications
Monitoring of
- Transformer oil
- Lubrication oil
- Hydraulic oil
- Diesel fuel
Key Features
- Measurands: water activity aw, temperature T and water content x in ppm
- Oil temperature up to 100 °C (212 °F)
- Two 4…20 mA outputs
- Modbus RTU interface
- Pressure rating 20 bar
- G ½” ISO or ½” NPT process connection
- Outputs freely configurable by the user
- Oil temperature up to 100 °C (212 °F)
Technical Data
Measuring range 0...1 aw -40...80 °C (-40...176 °F) or -40...100 °C (-40...212 °F) | Accuracy ±0.02 aw ±0.2 °C (±0.36 °F) |
Outputs 2 x 4-20 mA / MODBUS RTU | Supply 10-28 V DC |
Related products
Downloads EE364
Support literature
Manual
EU Declaration of Conformity
FAQ
What is the difference between Modbus RTU and Modbud TCP/IP
The main difference is the hardware interface. The Modbus RTU protocol runs on RS485 Hardware while Modbus TCP/IP protocol on Ethernet hardware.
What is the calibration for a specific oil? What are specific oil parameter?
Calibration for a specific oil is a laboratory procedure for determining the saturation curve of the oil. The saturation curved is described by two parameters (A and B). These parameters are used for setting up the moisture in oil sensor for correct calculation of the water content (ppm) out of the measured temperature and water activity (aw).
How can I replace the sensing element of the E+E MIO sensor?
Sensor replacement is possible the EE360 models when option PC4 (pluggable probe). Please see the operation manual for detail and spare part code ordering guide.
How can I clean the sensing element of the E+E MIO sensor?
Please see the cleaning instructions.
Can the coating be used for E+E MIO sensors?
No, coating is not available for MIO sensors.
What kind of calibration certificate is included in the standard scope of supply of the E+E MIO sensors?
The standard scope of supply of E+E MIO sensorsincludes an inspection certificate according DIN EN 10204 – 3.1.
Are there different A, B parameters for mineral transformer oil from various manufacturers?
No, mineral transformer oils from different manufacturers are quite similar.. A and B parameters are the same.
What is the different between a 2 wire and 3 wire current output?
Transmitter with “two-wires” technology receive the power from the process, and the signal is carry on return wire (closed current loop). With technology “three-wires” the power supply is separate from current output: 2 wires are the power supply and third carry the signal.